۲۳۷
الملاحق
ملحق رقم (۳)
مواد الوسط المسامي في الاستشراب العمودي السائل (المعكوس)
Mode
Reversed-phase HPLC: general information and packing materials
General information
Compounds determined Low- and medium-polarity compounds
Comments
Reversed phase
Most widely used chromatographic mode in clinical analysis
Mobile-phase ion suppression or ion pairing is required for ionic or ionizable compounds
Commercially available packing materials
Bonded phase
n-Octadecyl (C18), n-octyl (C8)
(Most widely used reversed-phase functionality in clinical analysis)
n-Butyl (C4)
(Preferred functionality for protein separations; however, less stable in acidic mobile phases [such as TFA] than C8, C18)
Phenyl
(Preferred functionality for compounds containing an aromatic group or groups)
(These functional groups can be attached to the following supports: silica, polybutadiene modified/PHAM/alumina PS-DVB, PVA)
Silica modified with alkyl polysiloxanes, multifunctional alkyl silanes, alkyl PVA
Alumina modified with polybutadiene (with or without attached alkyl groups)
Polymer covered
Base support (native)
PS-DVB
Nonporous
Perfusion
PAM
Porous graphite
PS-DVB (native), PHAM silica (reversed-phase-functionalities)
PS-DVB (native and C18 functionality) (from PerSeptive Biosystems)
Restricted access media internal surface reversed phase (ISRP) (from Regis Chemical)
(Glycine-1-phenylalanine--phenylalanine [GFF] functionalities)
Semipermeable surface (SPS) (from Regis Chemical)
(C8, C18, and phenyl functionalities)
Shielded hydrophobic phase (SHP) (from Supelco)
(Phenyl functionality)
PS-DVB, Poly(styrenemethyl divinylbenzene); PAM, poly(alkylmethylacrylate); PHAM, poly(hydroxyalkyl methacrylate or acrylate); PVA, poly(vinyl alcohol); TFA, trifluoroacetic acid.