Radiobiological and Microdosimetric Characteristics of ...
565
metric technique and the most suitable for that particular problem. Its response (or the G value) does not vary in depth. It does not introduce any heterogeneity and can occupy a volume identical to that occupied by the biological systems.
RBE as
As indicated in Table II, no significant variation of a function of depht has been observed in experiments performed using FeSO4 as the reference dosimeter.
nor
Experiments on chromosome aberrations in Allium Cepa (which was proven to be a system very sensitive to a change in radiation quality) did not show any significant RBE varia- tion in depth in the first millimeters of the irradiated medium, i.e. between the "surface" and the maximum of the depth dose curve, for a 15 MeV electron beam Obser- vation of skin reactions in patients resulted in the same conclusion (15).
These radiobiological results were confirmed by clinical systematic studies, as well as by the general clinical impres-
sions (16).
III.
COMPARISON OF MICRODOSIMETRIC CHARACTERISTICS OF HIGH-
ENERGY ELECTRON BEAMS, HIGH-ENERGY PHOTON BEAMS AND
CONVENTIONAL X-RAY BEAMS.
There is not any convincing physical reason that the clinical RBE could significantly differ between high-energy electrons and high-energy photons, above 1 MeV.
The degradated electron spectra, for the two radiations, are different only in their high-energy range, where the stopping power varies slowly with particle energy; on the other hand, the electron spectra are similar for both radia- tions in the low energy range, which is the most important for the RBE. Calculating of average LET were made in the early days of electrons therapy (5). However, more recently, direct measurements have shown that the distribution of the microdo- simetric characteristics are quite similar between high-energy electrons and high-energy photons(18) (20). Figure 1 compares the single event specific energy dose distributions for 39 MeV and 7.5 MeV electrons, 42 MV X-rays and 60Co y-rays(17). These distributions are quite similar. Figure 2 compares the single event specific energy dose distributions as a function of depth in a 15 MeV electron beam; again no important variation is observed. The data are summarized in Table III where the mean lineal energies for different beam qualities are compa- red. Similar conclusion can be drawn from this table. By