RELATIVE TOTAL DOSE
Figure 5
1.10
The Neutron Therapy Program at the Cyclotron “Cyclone"...
- 6
- 3
1.00-
0.90-
0
+3
+6
-1.10
PARALLEL OPPOSED FIELD
10 cm x 10 cm
patient thickness: 12 cm
T
- 6
- 3
T
0
60Co
d(50)-Be
-1.00
p(65)-Be
18 MV
+3
DISTANCE FROM MIDPOINT/CM
+6
-0.90
VD018208
: Variation of the total absorbed dose (D. ) along the central axis of two parallel opposed fields, for d(50)-Be and p(75)-Be, 10 cm x 10 cm, neutron beams. The patient thickness is supported to be The curves corresponding to60Co and 18 MV
12 cm.
X-rays are added for comparison.
(from Vynckier and Coll., in press).
213